• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0706-NTW
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
All available information was investigated, and the reported torn clip cover was confirmed.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the torn clip cover appears to be related to procedural conditions (interaction with anatomy).There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
It was reported on (b)(6) 2025 that the patient presented with grade 3 functional mitral regurgitation (mr), thick septum, previous cardiac surgery and restricted posterior leaflet for a mitraclip procedure.During the procedure, difficulty was encountered while performing transeptal puncture.A bulge was observed on the septum possibly due to previous cardiac surgery.An attempt was made to advance steerable guide catheter(sgc) into septum and resistance was felt.The clip was inserted into sgc and was observed under fluoroscopy and echo guidance.It was noted that sgc and clip were not inside the septum.The clip and sgc were removed from anatomy.Upon removal, visual inspection revealed damage to the polyester covering of the clip.A new mitraclip ntw device was prepared.An intra-aortic balloon pump was inserted to facilitate septal dilation.The sgc was reintroduced into the anatomy, and the ntw clip was successfully deployed.The mr was reduced to grade <1 post procedure.There was no clinically significant delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Common Device Name
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key23557299
Report Number2135147-2025-06655
Device Sequence Number13218753
Product Code NKM
UDI-Device Identifier05415067037435
UDI-Public(01)05415067037435(17)260224(10)50225R1014
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/24/2026
Device Model NumberCDS0706-NTW
Device Catalogue NumberCDS0706-NTW
Device Lot Number50225R1014
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2025
Initial Report FDA Received Date11/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/25/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
Patient Weight86 KG
-
-