The device was received for analysis.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were reviewed and all production steps were performed accordingly.In particular, the final acceptance test proved the device functions to be as specified.The memory content of the device was inspected.The device data documented a loss of signal after july 24, 2025.In a next step the device was opened to inspect the inner assembly and to check the functionality of the electronic module.During the analysis, it was revealed that the electronic module was damaged.The damage symptoms clearly indicate a temporary high current, probably precipitated by the reported cardioversion.In general, the device is protected against the high voltage discharge during an external cardioversion, but depending on the position of the external cardioversion electrodes and individual patient circumstances a damage of the electronic module cannot be excluded.In conclusion, the analysis of the inner assembly of the device revealed that the electronic module was damaged due to a temporary high current, which was probably precipitated by the external cardioversion.There was no indication of a material or manufacturing problem.
|