| Model Number |
TJF-150 |
| Medical Device Problem Codes |
Crack (1135); Material Discolored (1170); Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.The legal manufacturer reviewed the customer provided cleaning disinfection and sterilization (cds) processes where no obvious deviations from instructions for use (ifu) were identified.A root cause could not be identified.Based on the results of the investigation, most probable cause for the events discoloration inside light guide lens, the adhesive around the light guide lens had chipped was traced to component failure.A root cause could not be identified.Based on the results of the investigation, it is likely the most probable cause of the forceps elevator wire had contained foreign objects could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the duodenoscope exhibited discoloration inside light guide lens, adhesive around the light guide lens had chipped and forceps elevator wire had contained foreign objects.There was no patient involvement.
|
| |
|