Catalog Number 130790036
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Unspecified Infection (1930); Pain (1994); Unspecified Tissue Injury (4559)
Date of Event 10/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent the removal of delta xtend shoulder due to pain and suspected infection.Surgeon requested the position of the glenosphere and that the patient was 'notching'.There was also a fractured acromium.This suggested the deltoid was under severe tension when implant inserted.The delta xtend was removed and a djo altivate reverse shoulder was implanted.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: removal of delta xtend shoulder due to patient pain and suspected infection.Surgeon requested the position of the glenosphere and that the patient was 'notching'.There was also a fractured acromium.This suggested the deltoid was under severe tension when implant inserted.On this day [date] the delta xtend was removed and a djo altivate reverse shoulder was implanted.The product was not returned to depuy synthes; however, photos were provided for review.See attachment 20250908_055301578_ios.Heic, screenshot 2025-10-29 at 12.24.07.Png, img_0236.Jpeg, (b)(4) images (1), implant stickers (b)(4).The photo investigation revealed nothing indicative of a device nonconformance.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the dxtend screw lock d4.5x36mm would have not contributed to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: d4 (expiration date), d10 (concomitant), h3.Corrected: d4 (primary udi}.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X36MM
Common Device Name
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23560195
Report Number1818910-2025-19968
Device Sequence Number10108169
Product Code HSD
UDI-Device Identifier10603295027850
UDI-Public(01)10603295027850(17)280229(10)5519038
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K192855
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number130790036
Device Lot Number5519038
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/29/2025
Supplement Date Received by Manufacturer01/26/2026
Initial Report FDA Received Date11/14/2025
Supplement Report FDA Received Date01/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DXTEND GLENOSPHERE STD D42MM; DXTEND HIGHMO PE CUP D42 +3MM; DXTEND METAGLENE; DXTEND MOD EPI 2 ECC RIGHT HA; DXTEND MODULAR HUM STEM D12 HA; DXTEND SCREW LOCK D4.5X42MM; DXTEND SCREW NO LOCK D4.5X18MM; UNKNOWN SHOULDER NON-LOCKING SCREW
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
-
-