MEDTRONIC MEXICO S. DE R.L. DE CV AXIUM 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
QC-25-50-3D |
| Medical Device Problem Code |
Stretched (1601)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: as found condition: two axium devices were returned for analysis within a shipping box, within a sealed plastic biohazard pouch, within their opened axium inner pouch and within their introducer sheaths.The unknown micro catheter used in the event was not returned for analysis.Visual inspection/damage location details: (first axium device) no damages or irregularities were found with the introducer sheath.Liquid and dried blood were found within the introducer sheath and on the pusher/implant (figures e & g).The axium pusher was found bent/kinked between ~11.8cm and ~16.7cm from the proximal end (figure c) and between ~69.0cm and ~75.6cm from the proximal end (figure d).The axium implant coil was found stretched and damaged within the introducer sheath with the polypropylene filament unbroken (figure f).The implant was found broken within the introducer sheath with the polypropylene filament broken at the same location (figure h).The broken distal implant segment was found within a broken segment of an echelon micro catheter (figures i & j).The echelon segment measured about ~1.5cm with both ends found broken and the inner wire exposed and broken (figures i & j).(second axium device) the axium pusher was found broken near the break indicator with the release wire retracted (figure k).The axium pusher was found bent between ~86.0cm and ~88cm from the proximal end (figure l).The distal ~15.0cm of the pusher was also found bent/kinked (figure m).The coin was found retracted out of the lumen stop (figure n).No damages were found to the shield coil.The implant coil was already detached and found partially within the introducer sheath (figure o).The implant coil was found damaged within the introducer sheath.Testing/analysis: n/a conclusion: based on the device analysis and reported information, the customer¿s report of ¿coil stretch¿ was confirmed for the first axium device but not the second.However, both devices were found with the pusher bent/kinked and both implants were found damaged.Possible causes of coil stretching include, but not limited to, lack of hydration before procedure, user does not maintain continuous flush, tortuous anatomy, coil not retracted in a one-to-one motion with the implant pusher during repositioning, pushwire rotation, user advances the coil against resistance or incompatible catheter.Blood was found within the introducer sheath and on the device, potentially contributing towards resistance and coil stretching.It is possible that insufficient continuous flush was used, causing blood to backflow up the micro catheter and into the sheath.Customer reported vessel tortuosity as minimal, and devices were prepared per ifu.Therefore, the likely cause could not be determined.As the unknown micro catheter used in the event was not returned for analysis, any contribution of the micro catheter towards the coil stretching could not be determined.As the model and lot numbers were not reported, compatibility could not be assessed.The implant was found broken for the first axium device.It is likely the cause for the break is the same as the cause for coil stretching.The second device was found broken near the break indicator, with the release wire/coin retracted; detaching the implant as expected by the detachment subassemblies.The customer did not report attempting to detach the device.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that during transhepatic intrajugular portosystemic shunt (tips) surgery for vascular malformation embolization, the axium prime bare coil was selected for delivery, and the spring coil was stretched.It was replaced with a new medtronic coil to complete the procedure.No patient symptoms or complications were associated with this event.Additional information was received.The lesion was characterized as portal vein varicose veins.The timing of when the stretch was observed is unknown.No issues were reported that resulted in the damage found.The patient was undergoing surgery for treatment of varicose veins that were not located in the eloquent region.It was noted the patient's blood flow was high and vessel tortuosity was minimal.The reported device and any accessory devices were prepared as indicated in the instructions for use (ifu).
|
| |
|
Additional Manufacturer Narrative
|
|
Update to h3: as found condition: two axium devices were returned for analysis within a shipping box, within a sealed plastic biohazard pouch, within their opened axium inner pouch and within their introducer sheaths.The unknown micro catheter used in the event was not returned for analysis.Visual inspection/damage location details: (pli-10 and pli-30) no damages or irregularities were found with the introducer sheath.Liquid and dried blood were found within the introducer sheath and on the pusher/implant.The axium pusher was found bent/kinked between ~11.8cm and ~16.7cm from the proximal end and between ~69.0cm and ~75.6cm from the proximal end.The axium implant coil was found stretched and damaged within the introducer sheath with the polypropylene filament unbroken.The implant was found broken within the introducer sheath with the polypropylene filament broken at the same location.The broken distal implant segment was found within a broken segment of an echelon micro catheter.The echelon segment measured about ~1.5cm with both ends found broken and the inner wire exposed and broken.(pli-20) the axium pusher was found broken near the break indicator with the release wire retracted.The axium pusher was found bent between ~86.0cm and ~88cm from the proximal end.The distal ~15.0cm of the pusher was also found bent/kinked.The coin was found retracted out of the lumen stop.No damages were found to the shield coil.The implant coil was already detached and found partially within the introducer sheath.The implant coil was found damaged within the introducer sheath.Testing/analysis: n/a conclusion: based on the device analysis and reported information, the customer¿s report of ¿coil stretch¿ was confirmed for the first axium device but not the second.However, both devices were found with the pusher bent/kinked and both implants were found damaged.Possible causes of coil stretching include, but not limited to, lack of hydration before procedure, user does not maintain continuous flush, tortuous anatomy, coil not retracted in a one-to-one motion with the implant pusher during repositioning, pushwire rotation, user advances the coil against resistance or incompatible catheter.Blood was found within the introducer sheath and on the device, potentially contributing towards resistance and coil stretching.It is possible that insufficient continuous flush was used, causing blood to backflow up the micro catheter and into the sheath.Customer reported vessel tortuosity as minimal, and devices were prepared per ifu.Therefore, the likely cause could not be determined.As the unknown micro catheter used in the event was not returned for analysis, any contribution of the micro catheter towards the coil stretching could not be determined.As the model and lot numbers were not reported, compatibility could not be assessed.The implant was found broken for the first axium device.It is likely the cause for the break is the same as the cause for coil stretching.The second device was found broken near the break indicator, with the release wire/coin retracted; detaching the implant as expected by the detachment subassemblies.The customer did not report attempting to detach the device.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|