The product has not returned for evaluation.A radiographic image was provided which confirms the event.The locking mechanism has broken off, and the inferior screw has fractured and begun to backout.Multiple attempts have been made to obtain information.Based on the information provided, the root cause could not be determined.Alphatec spine is submitting this report to comply with the fda regulations 21 cfr 803.Alphatec spine has made reasonable efforts to provide as much relevant information as available.Any field that is left blank was not known at the time of this submission.If additional information and/or the implant are provided, a supplemental report will be filed accordingly.Labeling review: warnings/cautions: potential risks identified with the use of these fusion devices, which may require additional surgery, include device component failure, loss of fixation, pseudoarthrosis (i.E., non-union), fracture of the vertebra, neurological injury, and/or vascular or visceral injury.Possible adverse effects: initial or delayed loosening, bending, dislocation and/or breakage of device components.Postoperative management: the surgeon should instruct the patient regarding the amount and time frame after surgery of any weight bearing activity.The increased risk of bending, dislocation, and/or breakage of the implanted devices, as well as an undesired surgical result are consequences of any type of early or excessive weight bearing, vibratory motion, falls, jolts, or other movements preventing proper healing and/or fusion development.Implanted devices should be revised or removed if bent, dislocated, or broken.Immobilization should be considered in order to prevent bending, dislocation, or breakage of the implanted device in case of delayed, malunion, or nonunion of bone.Immobilization should continue until a complete bone fusion mass has developed and been confirmed.
|