• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SALTER LABS N/A; NASAL CANNULA (ADULT) HIGH FLOW WITH 7' (2.1 M) SUPPLY TUBE, FOR FLOWS UP TO 15

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SALTER LABS N/A; NASAL CANNULA (ADULT) HIGH FLOW WITH 7' (2.1 M) SUPPLY TUBE, FOR FLOWS UP TO 15 Back to Search Results
Model Number 1600HF-7
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Hypoxia (1918)
Date of Event 10/13/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 14 nov 2025 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.
 
Event or Problem Description
It was reported: one of the tubes connected to the nasal tip became detached, resulting in a significant drop in the amount of oxygen delivered to the patient(s).The issue was resolved by replacing the cannula(s).The patient(s) appeared to have suffered no lasting effects; however, slight hypoxaemia was noted.
 
Additional Manufacturer Narrative
The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 14 nov 2025 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as complaint - (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.The complaint of "nasal tip became detached" regarding part 1600hf-7-25 was confirmed.The root cause was not determined.A risk assessment was performed, and the ultimate risk was determined to be low which does not require escalation to the capa review board.There have been no other complaints regarding the same part and a similar issue within the 24 months preceding this reported event.A qc alert was created.A resolution letter was sent to the customer.Complaints will continue to be monitored for possible trends.
 
Event or Problem Description
It was reported: one of the tubes connected to the nasal tip became detached, resulting in a significant drop in the amount of oxygen delivered to the patient(s).The issue was resolved by replacing the cannula(s).The patient(s) appeared to have suffered no lasting effects; however, slight hypoxaemia was noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
N/A
Common Device Name
NASAL CANNULA (ADULT) HIGH FLOW WITH 7' (2.1 M) SUPPLY TUBE, FOR FLOWS UP TO 15
Manufacturer (Section D)
SALTER LABS
30 spur drive
el paso TX 79906
Manufacturer (Section G)
SALTER LABS DE MEXICO DE S.A. DE C.V.
blvd. independencia #2167 parq
ind. las ameri - hor sur cd ju
chihuahua mexico, mexico 32596
MX   32596
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key23563138
Report Number3004748541-2025-00104
Device Sequence Number10081804
Product Code CAT
UDI-Device Identifier00607411101625
UDI-Public(01)00607411101625(10)532690
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1600HF-7
Device Catalogue Number1600HF-7
Device Lot Number532690
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2025
Supplement Date Received by Manufacturer10/17/2025
Initial Report FDA Received Date11/14/2025
Supplement Report FDA Received Date01/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-