• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SA60AT
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Inflammation (1932)
Date of Event 09/26/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for analysis; the lens remains implanted.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A physician reported that following an intraocular lens (iol) implantation procedure, vitreous inflammation occurred within the second week after surgery.The patient was currently undergoing treatment.The procedure was completed successfully.Additional information was received stating lens was not explanted.There are four medical device reports associated with this report.This is 4 of 4.
 
Additional Manufacturer Narrative
Additional information was provided in h.3., h.6.And h.11.The iol was not returned.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.A non-company viscoelastic was indicated.The handpiece information was not provided.The product investigation could not identify a root cause for the reported complaint.The lens remains implanted.All iols (intraocular lens) are terminally sterilized with 100% ethylene oxide sterilization.Company uses an overkill sterilization approach, which greatly exceeds the minimum requirements for sterility.Critical parameters for sterilization cycles are recorded and retained for every sterilization load to demonstrate that the acceptance criteria for the sterilization process are met.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23563734
Report Number1119421-2025-02952
Device Sequence Number9764603
Product Code HQL
UDI-Device Identifier00380652555845
UDI-Public(01)00380652555845(17)300114(11)250116(10)15976446
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSA60AT
Device Catalogue NumberSA60AT.220
Device Lot Number15976446
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/21/2025
Supplement Date Received by Manufacturer11/24/2025
Initial Report FDA Received Date11/15/2025
Supplement Report FDA Received Date12/19/2025
Date Device Manufactured01/16/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III CARTRIDGE C; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C; NON COMPANY VISCOELASTIC
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-