Model Number DAQ916
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that pre-op, the entire nim setup does not work.Initially there was a problem with the cable from the stimulator ex tender, it had damaged pins and the stimulator was not working.The cable was replaced together with another stimulation box but the stimulator still did not work.And on the nim screen the error appeared: electro stimulator module failure: 0x00-st.There was no p atient involvement.
 
Additional Manufacturer Narrative
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: ceex20, serial/lot #: unknown, udi#: unknown, medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H6: previously applied code fdc d16 is no longer applicable.Applicable code is fdc d20.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: ceex20, serial/lot #: unknown, udi#: unknown.H6: previously applied code fdc d16 is no longer applicable.Applicable code is fdc d02.Additional information received shows that there is no information to reasonably suggest that the devices in this report may have caused or contributed to a death or serious injury or that the devices in this report has malfunctioned.Therefore, this event does not meet the reporting requirements in 21 cfr 803.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information received, there was no other issue from nim apart from the error message.And only one stimulator extender was used in this event, did not try with another one.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM-ECLIPSE® PREAMPLIFIER
Common Device Name
STIMULATOR, ELECTRICAL, EVOKED RESPONSE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key23566387
Report Number1045254-2025-02569
Device Sequence Number13226511
Product Code GWF
UDI-Device Identifier00643169378476
UDI-Public(01)00643169378476(11)210113
Combination Product (Y/N)N
PMA/510(K) Number
K050798
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDAQ916
Device Catalogue NumberDAQ916
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2025
Supplement Date Received by Manufacturer11/24/2025
Initial Report FDA Received Date11/17/2025
Supplement Report FDA Received Date12/16/2025
Date Device Manufactured01/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-