• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA CONST TASP SHIM CD 14MM; KNEE INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA CONST TASP SHIM CD 14MM; KNEE INSTRUMENT Back to Search Results
Catalog Number 42527900204
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Suggested component code: mechanical (g04) ¿ provisional.The customer has indicated product has been returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that instrument tap(s) were returned disassembled on a worn instrument claim form and that not all pieces were returned.Attempts to obtain additional information have been made; however, no more is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d1; d2; d4; g1; g3; g6; h1; h2; h3; h6; h9.The reported event can be confirmed.Visual examination of the returned tasp identified signs of use (scuffs, scratches, dings) and confirmed the complaint.One of the bearings and springs are disassembled / missing and not returned.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA CONST TASP SHIM CD 14MM
Common Device Name
KNEE INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23568265
Report Number0001822565-2025-04142
Device Sequence Number15919743
Product Code MBH
UDI-Device Identifier00889024245730
UDI-Public(01)00889024245730(10)63650570
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K123459
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number42527900204
Device Lot Number63650570
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2025
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2025
Supplement Date Received by Manufacturer04/22/2026
Initial Report FDA Received Date11/17/2025
Supplement Report FDA Received Date04/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/06/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-