• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC IMP,TSV,4.7,8,MTX,MG; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC IMP,TSV,4.7,8,MTX,MG; DENTAL IMPLANT Back to Search Results
Catalog Number TSVTWB8
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 11/04/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: patient weight unknown / not provided.G4: additional pma/510(k) number k101880.
 
Event or Problem Description
It was reported that the implant at tooth site #3 was removed due to infection.Discovered infection around implant, removed implant, curetted site and placed new implant.
 
Additional Manufacturer Narrative
Similar complaints for implant infection have been previously investigated.Visual and dimensional evaluations of the previously returned product have not identified or suggested manufacturing non-conformances.While non-conformances were identified for some lots during manufacturing record reviews, the documented disposition actions for each did not suggest the likely release of non-conforming product.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming product.Additionally, all device history record reviews verified that each implant was sterilized per procedure for every device.All complaint data used for the summary investigation was found to be conforming and did not meet capa/hhe/d/ or any further escalations.Therefore, there were no complaints which confirmed a manufacturing or design related issue that did or could cause or contribute to the reported event.Dhr review was completed for the subject lot number 1289373.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record op# (b)(4) was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number 1289373 for similar events and no other complaint was identified.As documented in the summary investigation, contributing factors for the reported event likely exist outside of zimvie control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigation, no malfunction occurred upon investigation.The reported event remains non-verifiable as it is a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for the infection of an implant have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,4.7,8,MTX,MG
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark  mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key23568372
Report Number0001038806-2025-03199
Device Sequence Number18623001
Product Code DZE
UDI-Device Identifier00889024020030
UDI-Public(01)00889024020030(17)291016(10)1289373
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K101977
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Catalogue NumberTSVTWB8
Device Lot Number1289373
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/17/2025
Supplement Report FDA Received Date12/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/17/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-