| Catalog Number |
FM-410310 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/01/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade ii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "implant is textured a type that has been associated with specific risks and have been recalled", "some cysts noted", "capsular contracture baker grade ii", "including severe food sensitivities", "significant fatigue".This record is for the left side.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6b, h6.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.¿ cyst(s): unable to observe through visual inspection as it is a physiological phenomenon.¿ fatigue: unable to observe through visual inspection as it is a physiological phenomenon.¿ other - medical: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (creases, wear abrasion) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "implant is textured a type that has been associated with specific risks and have been recalled", "some cysts noted", "capsular contracture baker grade ii", "including severe food sensitivities", "significant fatigue".This record is for the left side.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|