It was reported that a patient underwent a carotid endarterectomy on 2025 and suture was used.A highly experienced vascular surgeon, used article 8706 for carotid sutures.Unfortunately, they couldn't give me a specific date for the surgery.He's been complaining for months that the needles keep coming loose from the thread, sometimes even during removal from the wound support.All previous complaints have been inconclusive, which annoys him greatly.The doctor a highly experienced and accomplished vascular surgeon and has always been very satisfied with our sutures.There were no patient consequences reported.Additional information was requested.
|
|
Product complaint # (b)(4) date sent to the fda: 11/17/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested; the following was obtained: could you please indicate how many of these nine sutures involved had the needle detach from the thread during removal from the packaging? no could you please indicate how many of these nine sutures involved had the needle detach from the thread during passage through the tissue? no please provide the source or name of person providing answers to follow-up questions: op-(b)(6).No further information available.(b)(4) captures the initial report for ".The needles keep coming loose from the thread, sometimes even during removal from the wound dressing." bei der entnahme aus dem folienblister! when removing from the foil blister pack! (b)(4) captures the ambiguous multiple event report.What is the total number of procedures? no longer identifiable have any of these events been previously reported to ethicon? if so, could you please provide the corresponding reference number(s)? no, because all past complaints were always without any findings, which massively annoyed the very experienced and skilled chief physician.If this event occurred during multiple procedures, would you kindly create a product complaint for each event and provide the corresponding reference number(s)? not possible as no further information known and (b)(4) captured the ambiguous events when did the needles detach from the suture thread (removal from package / during handling prior to use on patient/ during passage through tissue / during tying / post-op)? partly during removal from the packaging, partly during passage through the tissue.Please confirm below, how many devices demonstrated the reported alleged deficiency? product code eh7223h - lot# 1053ru , 1 ea product code 8706h - lot# 101bhb , 9 ea please provide the product return status will send, sales rep is waiting for safety boxes provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq).Op-nurse (b)(6).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The following information was reported: the contact person is (b)(6).Surgical assistant.Was surgery delayed due to the reported event? yes, if yes, number of minutes: 20, was procedure successfully completed? yes, were fragments generated? yes, if yes, were they removed easily without additional intervention? yes, patient status/ outcome / consequences no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: no, is the patient part of a clinical study unknown, (b)(4), device property of none.
|