| Model Number |
1000D |
| Medical Device Problem Codes |
High impedance (1291); Incomplete or Inadequate Connection (4037)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient has been referred for lead revision due to current not being delivered.Internal data was later reviewed and confirmed that the patient is experiencing high impedance.No known relevant surgical intervention has occurred to date.No other relevant information has been received to date.
|
| |
|
Additional Manufacturer Narrative
|
|
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
|
| |
|
Event or Problem Description
|
|
Additional information received.It was later reported that the patient underwent pin re-insertion surgery and high impedance was resolved.No other relevant information has been received to date.
|
| |
|
Event or Problem Description
|
|
More information received from the provider.It was noted that x-rays were performed.It was also noted that the patient did not experience any trauma or manipulation around the vns after implant.X-rays were not made available for review.No other relevant information has been received to date.
|
| |
|