Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Nerve Damage (1979)
Date of Event 05/19/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: 110010247 g7 osseoti 4 hole shell 58mm g 67035482.110024465 g7 dual mobility liner 46mm g 66664855.110031013 28mm i.D.46mm o.D.Size g bearing 67051331.010001002 g7 screw 6.5mm x 45mm 6514808.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported a patient had an initial left total hip arthroplasty with competitor products followed by a revision for an unknown implant failure.During the procedure, the competitor stem was retained, and all other products were revised.The patient is now reporting significant nerve damage.No further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: a5, a6, b4, b5, d4 (expiration date, udi #), g3, g6, h2, h3, h6, h11.Corrected: h4.The reported event could not be confirmed due to lack of product/information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: patient had an initial left tha with depuy implants.Patient had a revision surgery.Patient reported icd coding corresponding to broken internal left hip prosthesis as an indication for revision.The depuy stem was retained, all other components of depuy (shell, liner and head) were revised.Zimmer biomet g7 shell, liner, bearing and screws were implanted, along with a depuy biolox head.No operative notes provided of the initial or revision surgery.Patient reported via a phone call to zb representative that they received mixed implants with zb and depuy components.The patient has significant nerve damage post-op.No medical records provided related to the issue.A definitive root cause cannot be determined.However, an off-label use was identified as zb implants were used in conjunction with competitor products.Zimmer biomet has not confirmed the compatibility for these combinations of devices.It is unknown if these off-label usages may have caused or contributed to the reported events.Please note per the ifu 87-6200-021-99 en rev.B it states "do not use dual mobility longevity and vivacit-e polyethylene hip bearings with components of any other system or manufacturer, unless authorized by zimmer biomet." and ifu 87-6203-367-99 en rev.C states, "zimmer biomet has not tested the safety or effectiveness of these devices for use in combination with non-zimmer biomet products or components.If surgeons elect to assemble and implant a construct that includes components not manufactured or distributed by zimmer biomet or its affiliates, they do so in reliance on their own clinical judgment and should so inform their patients".Both ifu's state " zimmer biomet or its affiliates are not liable for complications and/or failure that may arise from use of the device in circumstances outside of zimmer biomet or its affiliates control including, but not limited to, product selection and deviations from the device¿s indicated uses or surgical technique." if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCREW SELF-TAPPING 6.5 MM DIA. 30 MM LENGTH
Common Device Name
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23569569
Report Number0002648920-2025-00317
Device Sequence Number9924227
Product Code LPH
UDI-Device Identifier00889024119826
UDI-Public(01)00889024119826(17)340501(10)66673233
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K934765
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number00625006530
Device Lot Number66673233
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2025
Supplement Date Received by Manufacturer04/13/2026
Initial Report FDA Received Date11/17/2025
Supplement Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-