| Catalog Number |
08814848162 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of false-positive elecsys hbsag ii results for 1 patient sample on a cobas e 801 analytical unit.The initial result was 1.02 coi (reactive).The sample was repeated, and the results were 1.01 coi (reactive) and 0.998 coi (reactive).On (b)(6) 2025, confirmatory testing confirmed that the sample was non-reactive.A new sample was obtained from the patient, and the elecsys hbsag ii result was non-reactive.
|
| |
|
Additional Manufacturer Narrative
|
|
The analyzer's serial number is (b)(6).The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The qc and calibration were acceptable.A general reagent issue was excluded.The field service representative verified alignments, performed checks with acceptable results, and performed tests with acceptable results.No abnormalities were found.The investigation did not identify a product problem.
|
| |
|
Search Alerts/Recalls
|