| Model Number |
DDBB1D1 |
| Medical Device Problem Code |
Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
08/20/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right atrial (ra) lead exhibited undersensing triggering device classified false termination of atrial tachycardia/atrial fibrillation (at/af)event(s) at times.Additionally, it was noted that the patient has a history of inappropriate shocks from their implantable cardioverter defibrillator (icd).The ra lead and icd remain in use.No further patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|