MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER TREVISIO INTRAVASCULAR DELIVERY SYSTEM; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER TREVISIO INTRAVASCULAR DELIVERY SYSTEM; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 9-ATV08F45/80
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2025, a 8f and 9f amplatzer trevisio delivery systems were selected for a procedure.During preparation, issues were observed sequentially with both systems.The extension tubing were spraying and leaking below the locks.Pecifically, the extension tubing on each system exhibited leakage in the form of a saline spray, with an estimated volume of approximately 20 ml.No visible damage was noted on the tubing or other components.A pre-use inspection of both delivery systems was performed, and no abnormalities were identified at that timethe procedure was completed using a 12 f amplatzer trevisio delivery system.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.
 
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2025, a 8f and 9f amplatzer trevisio delivery systems were selected for a procedure.During procedure issues were observed sequentially with both systems.The extension tubing were spraying and leaking below the locks.Specifically, the extension tubing on each system exhibited leakage in the form of a saline spray, with an estimated volume of approximately 20 ml.Both delivery systems entered the patients anatomy.It was reported that they were considered as the issue.The extension tubing was reported to be the one with the issue.No visible damage was noted on the tubing or other components.A pre-use inspection of both delivery systems was performed, and no abnormalities were identified at that timethe procedure was completed using a 12 f amplatzer trevisio delivery system.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.
 
Additional Manufacturer Narrative
An event of fluid leak and fracture was reported.The device was returned to abbott for investigation and when analyzed, the extension tube was confirmed to be cracked.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all defined manufacturing specifications.There were no complaints associated with any other devices from the lot.The event appears to be related to a potential product quality issue, exception issue (b)(4) has been initiated on 31 january 2025 to further investigate this complaint.The investigation determined that the cracking observed in the female luer component of the amplatzer delivery system was due to excessive manual assembly force applied during manufacturing.This force introduced residual stress into the plastic material, which over time and under elevated temperatures led to delayed cracking.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER TREVISIO INTRAVASCULAR DELIVERY SYSTEM
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key23570967
Report Number2135147-2025-06690
Device Sequence Number18700325
Product Code DQY
UDI-Device Identifier05415067030924
UDI-Public(01)05415067030924(17)280331(10)10795659
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K190581
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9-ATV08F45/80
Device Catalogue Number9-ATV08F45/80
Device Lot Number10795659
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2025
Supplement Date Received by Manufacturer01/06/2026
Initial Report FDA Received Date11/17/2025
Supplement Report FDA Received Date01/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
-
-