| Model Number |
B1040-040 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation of a 4.00x40mm armada 35 balloon dilatation catheter, a leak was noted.There was no device use or patient involvement, and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection and functional testing performed on the returned device.The reported leak was not confirmed.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the reported information and the observations from the returned analysis, a definitive cause for the reported leak could not be determined.In this case, it may be possible that insufficient preparation and/or the sidearm and the syringe (device use for preparation) did not form a secure connection or had a loose connection resulting in the reported leak; however, a conclusive cause cannot be determined since the complaint could not be confirmed during return analysis.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|