| Catalog Number |
82250240S |
| Medical Device Problem Code |
Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information has been requested and if received, will be provided in the supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "distal drill interfered with the nail and could not drill into the nail easily.The physician could drill with difficulty while having the sleeve controlled by fingers.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "distal drill interfered with the nail and could not drill into the nail easily.The physician could drill with difficulty while having the sleeve controlled by fingers.".
|
| |
|
Search Alerts/Recalls
|