| Catalog Number |
V947H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 11/17/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.D4/g4: device is not distributed in the united states but is similar to device marketed in the usa.Therefore, (01) gtin is not available.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a spine procedure in (b)(6) 2025 and suture was used.Needle breakage during use on patient.No patient harm.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 12/17/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 pending evaluation of returned device.Additional information: d9, h3, h6.Additional information was requested, the following was obtained: product code: (4) x4407.Product lot/batch: 103a2r.Tracking#: (b)(4) (hamburg de).Received at cg labs on 12/1/2025.1.Could you please clarify if extra device belongs to this complaint? yes, the sales rep who picked up the devices oversea it a device has been received; however, it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 1/16/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.Corrected information: d1, d2a, d2b, d4, g1, g4.Additional information: d9, h3, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to eth for evaluation.Three unopened samples and one empty overwrap packet were received for analysis.To evaluate the condition of the returned samples, the packets were opened.The swage and attachment area were noted as expected.The tip and body of the needles were examined under magnification and no anomalies or issues related to needle breakage were found.The samples were tested by resistance test to needle and met the requirements.As part of eth quality process all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the devices were returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6.Corrected data: d2a, d2b, d1, d4, h4, g1, h11.Both product code and lot have been updated.D4/g4: device is not distributed in the united states but is similar to device marketed in the usa.Therefore, (01)gtin is not available.Additional information was requested, the following was obtained: product code: (4) x4407.Product lot/batch: 103a2r.Tracking# : (b)(4).Received at cg labs on 12/1/2025.1.Could you please clarify if extra device belongs to this complaint? yes the sales rep who picked up the devices oversea it.
|
| |
|
Search Alerts/Recalls
|
|