• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VANTIVE US HEALTHCARE LLC PRISMAFLEX SET; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VANTIVE US HEALTHCARE LLC PRISMAFLEX SET; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 106697
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual device was not available; however, photographs and a video of the sample were provided for evaluation.Visual inspection of the photos and video showed the reported leak from the set access post-pump pod.This component is manufactured by a vantive external supplier.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the pod defect was determined to be related to supplier manufacturing, and the issue is being further investigated.Nonconformance records have been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that an external blood leak was observed from the access pod of a prismaflex m100 set during continuous renal replacement therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SET
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
VANTIVE US HEALTHCARE LLC
round lake IL
Manufacturer (Section G)
VANTIVE HEALTHCARE - MEYZIEU
7, av. lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. il route 120
round lake, IL 60073
2242706717
MDR Report Key23576096
Report Number8010182-2025-00729
Device Sequence Number18524709
Product Code KDI
UDI-Device Identifier07332414064556
UDI-Public07332414064556
Combination Product (Y/N)N
Initial Reporter CountryVM
PMA/510(K) Number
K041005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number106697
Device Lot Number25A0042CB
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2025
Initial Report FDA Received Date11/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PRISMAFLEX MACHINE.
Patient SexUnknown
-
-