| Catalog Number |
MWF601S |
| Medical Device Problem Code |
Residue After Decontamination (2325)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of the investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Upon inspection of tray at warehouse, blood/bone found inside of instrument.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction - the fda registration number is (b)(6).The reported event could be confirmed, since the product was returned for evaluation and matches the alleged failure mode.The received device after decontamination shows the presence of contamination due to dry blood on top of the compactor surface.No presence of any bio matter into the grooves is observed.However, the device is extensively used as evident by wear, scratches, discoloration, abrasion marks and faded laser markings for the inclination angle.Overall damage indicates frequent usage.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material manufacturing or design related problems were found during the investigation.Based on the investigation, the root cause of the event was determined to be related to wear issues.The device in question has experienced extensive usage over time, which contributed significantly to the problem as it was manufactured in 2020.A scope of improvement is in progress to prevent the recurrence of the event.If more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
Upon inspection of tray at warehouse, blood/bone found inside of instrument.
|
| |
|
Search Alerts/Recalls
|