| Catalog Number |
650-1055 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
10/09/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: cat# 110024461 lot# 404810 g7 dual mobility liner 38mm c.Cat# 650-1064 lot# 3123237 cer option type 1 tpr sleve -6.G2: foreign - event occurred in japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately three years post-implantation, the patient underwent a hip revision due to the head and bearing disassociation, consistent with intraprosthetic dislocation.The head and bearings were revised.Attempts have been made and no further information has been provided.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h2, h3, h6, h11.The reported event could not be confirmed due to lack of product/information.No product was returned or pictures provided; a product evaluation could not be performed.Medical records were not provided.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|