| Model Number |
CL-914 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a c-section, the connection on the needle and thread broke while suturing.It was noted that three sutures from the same product and lot number was involved in the same event with the same patient and that the surgical time was extended. a new device from another manufacturer was used to resolve the issue.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: cl-914, cl-914 polysorb* 0 vio 90cm gs24 x36 (lot#a5b1689y); cl-914, cl-914 polysorb* 0 vio 90cm gs24 x36 (lot#a5b1689y); cl-914, cl-914 polysorb* 0 vio 90cm gs24 x36 (lot#a5b1689y) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6, h3 evaluation summary: medtronic conducted an investigation based upon all information received.Fifty-four of sutures representative samples were returned for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the connection on the needle and thread broke while suturing.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|