It was reported that the catheter included in the cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter tray was the incorrect length prior to a central venous catheter placement for medication administration and pressure monitoring during and after cardiac surgery procedure for an unknown patient.As the device was opened and prepared for the procedure, it was discovered that the hub was labeled 7fr 20cm, while the device packaging indicated the device was a 7fr 15cm kit.The length of the 7fr 20cm catheter was determined to be too long for the patient, so a new 7fr 15cm kit was opened to complete the procedure successfully.The catheter did not patient contact.The patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
|
Blank fields on this form indicate the information is unchanged, unknown, or unavailable.Correction: d9, h3 - the device was not received; h6 - annex a.Investigation ¿ evaluation: it was reported that the catheter included in the cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter tray was the incorrect length prior to a central venous catheter placement for medication administration and pressure monitoring during and after cardiac surgery procedure for an unknown patient.As the device was opened and prepared for the procedure, it was discovered that the hub was labeled 7fr 20cm, while the device packaging indicated the device was a 7fr 15cm kit.The length of the 7fr 20cm catheter was determined to be too long for the patient, so a new 7fr 15cm kit was opened to complete the procedure successfully.The catheter did not patient contact.The patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), and quality control procedures for the device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported lot found that work orders were swapped and the sets were reworked.A database search for complaints on the reported lot found one additional related complaint.There is evidence of additional non-conforming product in the field.The information provided upon review of the dmr and dhr indicates the devices were manufactured out of specification.However, appropriate measures have been initiated to address this failure mode; a non-conformance investigation was opened to further investigate the issue.Based on the information provided, no device return, and the results of the investigation, cook concluded the event to be related to a manufacturing and quality control deficiency.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|