• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW CVC SET: 2-LUMEN 7 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW CVC SET: 2-LUMEN 7 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE Back to Search Results
Catalog Number CS-27702-E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "(b)(6) 2025, the ars was found leak during use on the patient." the user changed to a new set to resolve the issue.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "(b)(6) 2025, the ars was found leak during use on the patient." the user changed to a new set to resolve the issue.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The customer returned one arrow raulerson syringe (ars) for analysis.The lidstock included all other components of the kit except the 18 ga introducer needle and pressure transduction probe.Signs of use, such as biological material, were observed inside the ars.Visual inspection of the ars did not reveal any anomalies or defects.The returned sample was functionally evaluated using the lab inventory 18-gauge introducer needle, per the instructions for use provided with this kit.The ifu states, "insert introducer needle with attached syringe or arrow raulerson syringe (where provided) into vein and aspirate".The ars was able to draw and aspirate water with and without the introducer needle attached.No leaks or excessive air buildup were observed.The module requirement document for raulerson syringe (amrq-000113 rev.03) was reviewed to determine requirements for air/water leakage.The document notes a deviation from iso 7886-1: "the freedom of air and liquid leakage past the piston requirement is design restrictive and is intended for an injection-intended syringe, not the ars.The opening in the center of the piston that allows passage of the inner cannula prohibits the ars from meeting the pressure and vacuum requirements as dictated by the standard.However, because the intended use of the ars is to allow aspiration of blood to ensure venous placement of the introducer needle and to aid in the insertion of the spring wire guide, the leakage requirements of a standard syringe are not applicable to the ars." a vacuum test was performed on the ars in order to verify that the internal valves within the plunger body were intact.With the plunger body at the bottom of the syringe, the tip of the ars was occluded, and the plunger was pulled back until it stopped.With the tip of the ars still occluded, the plunger was released and snapped back into a position = 1cc from the starting position.Therefore, the internal valves of the ars were functioning as intended.A device history record review was performed, and no relevant findings were identified.The ifu provided with the kit informs the user, "insert introducer needle or catheter/needle with attached syringe or arrow raulerson syringe (where provided) into vein and aspirate." the customer report of a leaking ars could not be confirmed through investigation of the returned sample.Functional testing of the returned ars revealed no obvious defects or anomalies the ars was able to draw and aspirate water with and without a lab inventory 18ga introducer needle and passed all relevant additional testing.Therefore, based on the sample received and the customer report, no problem was found with the returned device.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 2-LUMEN 7 FR X 20 CM
Common Device Name
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key23582309
Report Number3006425876-2025-01056
Device Sequence Number21215527
Product Code FOZ
UDI-Device Identifier00801902102362
UDI-Public(01)00801902102362(11)241007(17)260913(10)71F24K0130(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K900263
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date09/13/2026
Device Catalogue NumberCS-27702-E
Device Lot Number71F24K0130
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2025
Supplement Date Received by Manufacturer11/21/2025
Initial Report FDA Received Date11/18/2025
Supplement Report FDA Received Date12/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED.; NOT REPORTED.
Patient SexUnknown
-
-