• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SILK ROAD MEDICAL ENROUTE TRANSCAROTID STENT; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SILK ROAD MEDICAL ENROUTE TRANSCAROTID STENT; STENT, CAROTID Back to Search Results
Model Number SR-0940-CS
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code Ischemia Stroke (4418)
Date of Event 10/22/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced an embolic stroke, requiring additional surgical intervention.An enroute transcarotid stent was selected for use in a left sided transcarotid artery revascularization (tcar) procedure.After completion of the procedure, the patient experienced neurological deficit.Imaging performed revealed an embolic stroke in the left hemisphere and the physician noted that the proximal end of the second stent was placed in a dissection plane in the common carotid artery (cca).The physician performed a cea to explant the stents and patched the carotid artery.
 
Additional Manufacturer Narrative
A secondary mdr was reported under 2124215-2025-82875 as there are two products associated with the same event/patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENROUTE TRANSCAROTID STENT
Common Device Name
STENT, CAROTID
Manufacturer (Section D)
SILK ROAD MEDICAL
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer (Section G)
SILK ROAD MEDICAL INC.
27th ave north plymouth
minneapolis MN 55447
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key23582708
Report Number2124215-2025-82852
Device Sequence Number9305510
Product Code NIM
UDI-Device Identifier00811311020508
UDI-Public(01)00811311020508(17)271130(10)18399741
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSR-0940-CS
Device Catalogue NumberSR-0940-CS
Device Lot Number0018399741
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2025
Initial Report FDA Received Date11/18/2025
Date Device Manufactured12/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age87 YR
Patient SexFemale
Patient Weight80 KG
Patient RaceWhite
-
-