G2: foreign country - slovenia h3/h6: the product analysis was completed.The tip of the returned probe had been severely twisted.There were also impact marks at the back end causing fit issues with the mating tracker.Codes b01, c07, and d02 apply.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|