|
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b4, d9, g3, g6, h2, h3, h6, h11.The following sections were corrected: h4.Visual evaluation of the returned product and packaging confirmed that the sterile packaging exhibited damage consistent with thermal damage, however, sterility was not breached.The device history records were reviewed and no discrepancies were identified.The root cause is attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|