| Catalog Number |
256045 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D4.Medical device expiration date: unknown.H4.Device manufacture date: unknown.The information for the additional 510k is as follows: g4.Pma/510(k)#: k132259; k132692; k151291; k152870; k160161.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using the bd veritor¿ system for rapid detection of flu a+b clia-waived kit, an unspecified number of false positive results were obtained.In addition, it was noted that qc failed.The customer did not specify if the analyte in question was flu a or flu b.There were no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: this statement summarizes the investigation results regarding a complaint that alleges false positive and qc swab - failure when using kit flu a+b 30 test physician veritor (material #: 256045), batch number unknown.The customer reported that they were receiving false positive results and qc failures.Bd quality performs a systematic approach to investigate all complaints.This approach involves review of manufacturing batch history records, visual inspection of retention samples, and visual inspection of customer returned samples, if applicable.Batch history record (bhr) review and retain sample testing were not performed as there was no batch number provided.No photos or samples were received; therefore, return sample analysis could not be performed.The reported issue was unable to be confirmed.A trend analysis for false positive and qc swab - failure was conducted, no adverse trend was identified.There were no corrective actions taken at this time.
|
| |
|
Event or Problem Description
|
|
It was reported while using the bd veritor¿ system for rapid detection of flu a+b clia-waived kit, an unspecified number of false positive results were obtained.In addition, it was noted that qc failed.The customer did not specify if the analyte in question was flu a or flu b.There were no health impact or consequences reported.
|
| |
|