• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THINK SURGICAL, INC TMINI MINIATURE ROBOTIC SYSTEM; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THINK SURGICAL, INC TMINI MINIATURE ROBOTIC SYSTEM; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 109343
Medical Device Problem Code Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2025
Type of Reportable Event Malfunction
Event or Problem Description
The user noticed pieces falling out of the tmini head during pre-op set-up of the product.Despite this, the tmini robot was used successfully.
 
Additional Manufacturer Narrative
Upon inspection of the returned tmini robot, think surgical, inc.Confirmed that one of the two the nut tubes was broken and the dowel pin was missing.The tmini functioned as intended as only one of the nut tubes was broken.Think surgical reviewed the device history record and confirmed there was no evidence of damage to the nut tube during production.There was no injury to patient, and the surgery was completed without complication using the tmini robot.However, think surgical is reporting this incident because if the dowel pin detached during the procedure, it could pose a risk of entering the surgical incision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TMINI MINIATURE ROBOTIC SYSTEM
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
THINK SURGICAL, INC
47201 lakeview blvd
fremont CA 94538
Manufacturer (Section G)
THINK SURGICAL, INC
47201 lakeview blvd
fremont CA 94538
Manufacturer Contact
denise duchene
47201 lakeview blvd
fremont, CA 94538
8604815113
MDR Report Key23587185
Report Number3000719653-2025-00018
Device Sequence Number15767871
Product Code OLO
UDI-Device IdentifierM9461093430
UDI-PublicM9461093430
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K252544
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number109343
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2025
Initial Report FDA Received Date11/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-