• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 550; CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 550; CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS Back to Search Results
Model Number 550AHCT
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use of the device for cardiopulmonary bypass (cpb), the blood parameter monitor (bpm) potassium (k+) reading was falling to 1.3 even though it was just calibrated and the k+ code entered.As mitigation, blood gas values from the blood gas analyzer were relied upon during the case.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Additional Manufacturer Narrative
Updated block: h6.The reported complaint was not verifiable since the product was not returned for evaluation or testing.Multiple diligence attempts for part return were unsuccessful.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 550
Common Device Name
CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key23587438
Report Number1828100-2025-00291
Device Sequence Number9178588
Product Code DRY
UDI-Device Identifier00886799001790
UDI-Public(01)00886799001790(11)220112
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K182110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number550AHCT
Device Catalogue Number550AHCT
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2025
Supplement Date Received by Manufacturer01/15/2026
Initial Report FDA Received Date11/18/2025
Supplement Report FDA Received Date02/05/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/12/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BLOOD GAS ANALYZER.
Patient SexUnknown
-
-