| Catalog Number |
M0035421030 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that resistance was felt while advancing subject coil and it got prematurely detached in the microcatheter.Physician pulled out coil and microcatheter together from patients' body.The procedure was completed successfully without any clinical consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The subject coil was not returned for analysis; therefore, physical as well as a functional testing could not be performed.The reported event is covered in the device direction for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The subject coil has not been returned for analysis.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned.
|
| |
|
Event or Problem Description
|
|
It was reported that resistance was felt while advancing subject coil and it got prematurely detached in the microcatheter.Physician pulled out coil and microcatheter together from patients' body.The procedure was completed successfully without any clinical consequences to the patient.
|
| |
|
Search Alerts/Recalls
|