BIOSENSE WEBSTER INC QDOT MICRO; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
| Catalog Number |
D139401 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot micro and an irrigation issue occurred during use on the patient.It was reported that after ablation of pulmonary vein isolation (pvi) with qmode+, a switch to cavotricuspid ishmus (cti) ablation with qmode was not possible as the catheter seems to be clogged and would not ablate.There was no patient consequence.There was no error on pump but, temperature and impedance rises rapidly on catheter during ablation and ablation session shuts off.The catheter was removed from patient and they tried flushing it, but no water came out.The catheter was replaced and issue resolved.There were no issues with flow rate change at the start of ablation.The correct catheter settings were selected on the generator.
|
| |
|
Additional Manufacturer Narrative
|
|
On 12-nov-2025, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot micro and an irrigation issue occurred during use on the patient.Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and irrigation test of the returned device were performed.Visual inspection revealed no damage or anomalies on the device.An irrigation test was performed, and the device was irrigating correctly, no obstructed holes were observed.No irrigation issues were observed.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The irrigation issue reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the use of the device that may have affected its performance.The instructions for use (ifu) contain the following information: purge the catheter and irrigation tubing with heparinized normal saline prior to insertion of the catheter inside the patient body as part of johnson & johnson medtech's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
| |
|