• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC. ARTIC-L¿ AND ARTIC-XL¿ SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC. ARTIC-L¿ AND ARTIC-XL¿ SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, Back to Search Results
Model Number 5330008
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a device used for spinal therapy.It was reported that the inserter experienced instrument breakage when the inserter broke and the pins in the inserters were backing out.The loose piece was identified before use, and no patient was involved.No patient complications have been reported as a result of this event.Additional information was received from the manufacturer representative that for the first inserter also, the pin was backing out making grip handle loose.
 
Additional Manufacturer Narrative
H3: product analysis : product# 5330008; lot # cn17g006 visual and optical inspection confirmed the laser weld on one side of the handle has broken.The damage appears to be from overload placed on the instrument during use medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIC-L¿ AND ARTIC-XL¿ SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key23590138
Report Number3003120897-2025-00590
Device Sequence Number21313353
Product Code MAX
UDI-Device Identifier00763000085902
UDI-Public(01)00763000085902(11)190731(10)CN17G006
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K171689
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5330008
Device Catalogue Number5330008
Device Lot NumberCN17G006
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2025
Initial Report FDA Received Date11/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-