RESPIRONICS, INC. REMSTARAUTOA-FLEX; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number AH560HBT
Medical Device Problem Codes Contamination (1120); Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted as part of a batch submission of complaints deemed reportable following record correction and remediation efforts.
 
Event or Problem Description
A device was returned to a third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the third-party service center, the device was visually inspected and found evidence of foam degradation.During the evaluation, secondary findings, dust was observed.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the third-party service center.There was 16 error code found.The third-party service center concludes that they could confirm the customer's allegation and there were visible foam degradation and unit got corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTARAUTOA-FLEX
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key23590966
Report Number2518422-2025-052786
Device Sequence Number21383921
Product Code BZD
UDI-Device Identifier00606959023925
UDI-Public(01)00606959023925
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation 505
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 11/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberAH560HBT
Device Catalogue NumberAH560HBT
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2024
Initial Report FDA Received Date11/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/15/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1