• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US INFLOW TUBING FMS VUE 24PK; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US INFLOW TUBING FMS VUE 24PK; ARTHROSCOPE Back to Search Results
Catalog Number 284508
Medical Device Problem Codes Crack (1135); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event or Problem Description
It was reported that during an unspecified surgical procedure a hole was noticed in the inflow tubing fms vue pump, which caused the hose to open, causing a large leak of saline solution.It was reported that another device was used to complete the surgery.There were no reported adverse consequences to the patient.The exact date of the event was not reported, however, it was reported that the event occurred in 2025.No additional information could be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFLOW TUBING FMS VUE 24PK
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key23592415
Report Number1221934-2025-04462
Device Sequence Number9924523
Product Code HRX
UDI-Device Identifier10886705023066
UDI-Public(01)10886705023066(10)3012120(17)260730
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K951843
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/30/2026
Device Catalogue Number284508
Device Lot Number3012120
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2025
Initial Report FDA Received Date11/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-