(b)(4) d10: 611777510 z1 starter broach unknown , 611889000 z1 broach sz 0 unknown , 611889010 z1 broach sz 1 unknown.Suggested component code- mechanical (g04): stem.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: a1, a3, a5, b1, b2, b4, b5, g3, g6, h1, h2, h3, h6, h10, h11.Visual examination of the returned product identified the size engraving is conforming for all 3 broaches.The stem has scratches, indentations, and gouges on the neck and trunnion surfaces.There are also multiple chips to the visible ha coating on the distal tip.All 3 broaches have scratches around the post and nicks on the cutting teeth and broach connection end.No other damage was noted during visual evaluation.All devices where conforming to specifications where measured.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined for the stem causing the fracture upon implantation.No problem was found with the broaches.They came back and measured within specification, with minor damage noted to the cutting edges.The surgical technique was followed and the surgeon broached sequentially up to the size of the definitive implant.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|