| Model Number |
CCA0T0 |
| Medical Device Problem Code |
Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A non-healthcare professional reported that during an intraocular lens procedure, iol haptics were not properly ejected and removed from wound before insertion with no patient harm.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3., h.6.And h.11.The product was returned for analysis and the reported complaint could not be observed.Iol was not returned.Only the device was returned.The device was returned loose in the carton.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been fully advanced outside the nozzle.No damage observed.Iol was not returned.The product investigation could not identify the root cause for the reported complaint "iol haptics were not properly ejected" as the lens was not returned.Only the device was returned for evaluation.Due to this, we are unable to confirm the lens and the plunger position for advancement and determine a root cause.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|