|
Bd technical support troubleshoot with customer over the phone.A device history record, complaint history review, and risk review on failure modes are performed on each device with reportable malfunction(s) along with other methods of investigation as coded in section h6 of this mdr report.Root cause: the probable cause of the reported issue was that the 8300 did not detect the luer installed, which was not identified during the customer call.The tech support recommended replacing the luer connector.Provided part number 49000938 board, oridion module assembly or sending in device for repair.Per 803.52(f)(11)(iii) the information provided represents all of the known information at this time.The complainant or reporter was unable or unwilling to provide any further patient, product, or procedural details to the manufacturer.
|