Radiometer reference: (b)(4).According to the complaint, five patient samples measured on the abl800 flex analyzer (serial no.(b)(6) were recalled on 2025-03-18.At 13:36, the analyzer was placed in stop mode without any documented action, replacement, or indication of the reason.-calibration was confirmed as ok at 11:34.-creatinine results were acceptable at 13:22 but not at 14:00, when the first patient sample showed an abnormally low creatinine value.-at 14:04, another sample was analyzed, returning "?" for all parameters.No instrument alarm was triggered after this event.-at 14:31, the equalis control was analyzed, yielding a result of 45 mol/l compared to the method group result of 128 mol/l.The deviation was so significant that the value was excluded from statistical analysis.There was no harm to patients.
|