• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RADIOMETER MEDICAL APS ABL800 FLEX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RADIOMETER MEDICAL APS ABL800 FLEX Back to Search Results
Model Number 393-801
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
Radiometer reference: (b)(4).According to the complaint, five patient samples measured on the abl800 flex analyzer (serial no.(b)(6) were recalled on 2025-03-18.At 13:36, the analyzer was placed in stop mode without any documented action, replacement, or indication of the reason.-calibration was confirmed as ok at 11:34.-creatinine results were acceptable at 13:22 but not at 14:00, when the first patient sample showed an abnormally low creatinine value.-at 14:04, another sample was analyzed, returning "?" for all parameters.No instrument alarm was triggered after this event.-at 14:31, the equalis control was analyzed, yielding a result of 45 mol/l compared to the method group result of 128 mol/l.The deviation was so significant that the value was excluded from statistical analysis.There was no harm to patients.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABL800 FLEX
Common Device Name
ABL800 FLEX
Manufacturer (Section D)
RADIOMETER MEDICAL APS
aakandevej 21
broenshoej, 2700
DA  2700
Manufacturer (Section G)
RADIOMETER MEDICAL APS
aakandevej 21
broenshoej, 2700
DA   2700
Manufacturer Contact
tom engdahl
aakandevej 21
broenshoej, 2700
DA   2700
MDR Report Key23593118
Report Number3002807968-2025-00134
Device Sequence Number21364636
Product Code CHL
UDI-Device Identifier05700693938011
UDI-Public(01)05700693938011(10)754R2731N001
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K041874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number393-801
Device Catalogue Number393-801
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2025
Initial Report FDA Received Date11/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/21/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-