OLYMPUS WINTER & IBE GMBH VIDEO TELESCOPE "ENDOEYE", 10 MM, 30°, NTSC, HD COMPATIBLE, AUTOCLAVABLE; RIGID VIDEO LAPAROSCOPE
|
Back to Search Results |
|
| Model Number |
WA50013A |
| Medical Device Problem Codes |
Break (1069); Poor Quality Image (1408); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the complaint is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the rigid video laparoscope exhibited failure of light source (light transmission issue), abnormal image, component failure of light guide bundle and electronics.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|
|
|