| Model Number |
2120 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: 2025 was the reported year of the event but the month/day was not reported.H3: the device was received for evaluation.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.Visual and functional tests were performed, and a malfunctioning downstream occlusion (dso) sensor was found.The dso sensor was replaced to fix the issue.
|
| |
|
Event or Problem Description
|
|
The device was returned due to a cassette sensor error.The results of the investigation found that the device's downstream occlusion (dso) sensor was malfunctioning.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|