Model Number 2120
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3: 2025 was the reported year of the event but the month/day was not reported.H3: the device was received for evaluation.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.Visual and functional tests were performed, and a malfunctioning downstream occlusion (dso) sensor was found.The dso sensor was replaced to fix the issue.
 
Event or Problem Description
The device was returned due to a cassette sensor error.The results of the investigation found that the device's downstream occlusion (dso) sensor was malfunctioning.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-SOLIS VIP AMBULATORY INFUSION PUMP
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
ICU MEDICAL, INC.
3350 granada avenue north
oakdale MN 55128
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key23593583
Report Number3012307300-2025-12590
Device Sequence Number15997102
Product Code FRN
UDI-Device Identifier15019517149999
UDI-Public(01)15019517149999(11)190502
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111275
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2120
Device Catalogue Number21-2120-0104-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2025
Initial Report FDA Received Date11/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-