| Catalog Number |
0165L18 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the foley catheter balloon burst during the pre-test.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter name:(b)(6).The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter balloon burst during the pre-test.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial reporter's facility name: (b)(6) hospital.The reported event was confirmed ¿ manufacturing related.The product had caused the reported failure.Sample was evaluated and observed there was a pinhole on sac body that allowed water to leak out.A potential root cause for this failure could be due to bubble during dipping process.The labeling and instructions for use were found to be adequate.A review of the manufacturing process indicates that the process is capable of producing of this type of defect.However, current in-process controls per pfmea are adequate to identify the defect or verify the product integrity.Corrections: d, e, f, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|