Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Product complaint # (b)(4).Investigation summary: according to the information received, "could not be implanted in place.It was reported that during the surgery, the implant could not implanted into the tibial plateau.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided." the product was not returned to depuy synthes, however photos were provided for review.Photos were provided for review, however the photos only show labels with part and lot number and no other observation pertaining to the nature of the reported event could be identified.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was not confirmed as the photographs provided contained insufficient evidence of the reported event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device product code 151640405 lot number m83n32, and no non-conformances / manufacturing irregularities were identified during manufacturing.Device history review: a manufacturing record evaluation was performed for the finished device product code 151640405 lot number m83n32, and no non-conformances / manufacturing irregularities were identified during manufacturing.
|