INSPIRE MEDICAL SYSTEMS INSPIRE MODEL 4323 RESPIRATORY SENSING LEAD Back to Search Results
Model Number 4323
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code Pain (1994)
Date of Event 11/06/2025
Type of Reportable Event Serious Injury
Event or Problem Description
During a routine follow-up examination, physician observed a device sensing issue.The patient reported severe, painful shocking sensations deeper in the chest, preventing them from lying down.Imaging was performed, and the distal tip of the sensing lead was suspected of separating from the lead body.Physician brought the patient in for surgery and confirmed the distal sensor tip had separated from the lead body and migrated.The physician consulted with a cardiothoracic surgeon and deemed retrieval was a higher risk to the patient, so the sensor was abandoned.The physician removed the sense lead body and implanted a newer ipg model, which does not require a sensing lead, with no further issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRE MODEL 4323 RESPIRATORY SENSING LEAD
Common Device Name
INSPIRE MODEL 4323 RESPIRATORY SENSING LEAD
Manufacturer (Section D)
INSPIRE MEDICAL SYSTEMS
5500 wayzatablvd.
suite 1600
golden valley MN 55416
Manufacturer (Section G)
INSPIRE MEDICAL SYSTEMS
5500 wayzata blvd.
suite 1600
golden valley MN 55416
Manufacturer Contact
vang yang
5500 wayzata blvd.
suite 1600
golden valley, MN 55416
7632057970
MDR Report Key23596407
Report Number3007666314-2025-00781
Device Sequence Number10291253
Product Code MNQ
UDI-Device Identifier10855728005496
UDI-Public10855728005496
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P130008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4323
Device Catalogue Number900-002-009
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2025
Initial Report FDA Received Date11/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-