| Model Number |
3156 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
10/22/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.Additional components potentially involved in the event include: common device name: scs lead, model: 3156, udi: (b)(4), serial: n/a, batch:4173940.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced ineffective therapy due to high impedances.As a result, surgical intervention may be undertaken to address the issue.It is unknown which leads have the high impedances.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
Additional information indicates that after a routine ipg replacement, effective therapy was restored postoperatively and intervention is no longer planned to replace the leads.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information indicates that after a routine ipg replacement, effective therapy was restored postoperatively and intervention is no longer planned to replace the leads.This device is no longer considered reportable.
|
| |
|