| Catalog Number |
367985 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There were multiple 510k numbers reported to be involved.The information is as follows: g.4.Pma / 510(k)#: k230855.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
Report 2 of 2: it was reported while using bd vacutainer® sst¿, an unspecified number of tubes burst during centrifugation.No adverse impact was reported regarding the patient or user.
|
| |
|
Event or Problem Description
|
|
Report 2 of 2: it was reported while using bd vacutainer® sst¿, an unspecified number of tubes burst during centrifugation.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received 3 photos for investigation.The photos show lot# 5189432 after collection with a crack on the sidewall.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the lot 5189432, for damaged/cracked component.This complaint has not been confirmed for molding defects.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|