• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIDEX CORPORATION G-PROBE RFID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIDEX CORPORATION G-PROBE RFID Back to Search Results
Lot Number 901431
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Eye Injury (1845)
Date of Event 07/15/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to iridex that while using the g-probe rfid, the patient experienced scleral burn.The reported settings used for treatment were consistent with the g-probe ifu (1500 mw for 3.5 seconds).During the procedure, burns were observed, the power was then decreased to 1400 mw, 1300 mw with scleral burns, and the probe was inspected and debris were not observed.No permanent impairment was reported to iridex, and no medical intervention associated with the burn was reported to iridex.The probe is disposable and was not returned to iridex for further investigation.Scleral burns are a known potential harm associated with the use of laser delivery in ophthalmic indications.Typically, scleral burns resolve within 24 to 48 hours.Therefore, unless either medical intervention or permanent impairment are reported, scleral burns are not deemed a serious adverse event, because scleral burns typically: do not result in life-threatening illness or injury, do not result in permanent impairment of a body structure or a body function, do not require hospitalization or prolongation of patient hospitalization, do not require medical or surgical intervention to prevent life-threatening illness or injury or permanent impairment to a body structure or a body function.The laser console used with the probe was the cyclo g6 laser console.The laser console was returned to iridex for further investigation and evaluation.As per the servicing carried out by service technicians, no problem was found with the console and it was returned to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G-PROBE RFID
Common Device Name
G-PROBE RFID
Manufacturer (Section D)
IRIDEX CORPORATION
1212 terra bella avenue
mountain view CA 94043 1824
Manufacturer (Section G)
IRIDEX CORPORATION
1212 terra bella avenue
mountain view CA 94043 1824
Manufacturer Contact
bill hyatt
1212 terra bella avenue
mountain view, CA 94043-1824
MDR Report Key23597920
Report Number2939653-2025-00005
Device Sequence Number21166469
Product Code GEX
UDI-Device Identifier00813125015237
UDI-Public00813125015237
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K143154
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Lot Number901431
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2025
Initial Report FDA Received Date11/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age62 YR
Patient SexMale
Patient EthnicityHispanic
-
-